                    Number
                 
             | 
            
                 
                    Technical projects
                 
             | 
            
                 
                    Specification parameters
                 
             | 
        
        
            | 
                 
                    1.
                 
             | 
            
                 
                    Temperature limit range:
                 
             | 
            
                 
                    -70 ℃ to 100 ℃ (no-load).
                 
             | 
        
        
            | 
                 
                    2.
                 
             | 
            
                 
                    Temperature control range:
                 
             | 
            
                 
                    -40 ℃ to 85 ℃ (standard load).
                 
             | 
        
        
            | 
                 
                    3.
                 
             | 
            
                 
                    Temperature fluctuation:
                 
             | 
            
                 
                    ≤ ± 0.5 ℃.
                 
             | 
        
        
            | 
                 
                    4.
                 
             | 
            
                 
                    Temperature uniformity:
                 
             | 
            
                 
                    ± 2.0 ℃.
                 
             | 
        
        
            | 
                 
                    5.
                 
             | 
            
                 
                    Temperature deviation:
                 
             | 
            
                 
                    ≤ ± 2.0 ℃.
                 
             | 
        
        
            | 
                 
                    6.
                 
             | 
            
                 
                    Heating rate:
                 
             | 
            
                 
                    Approximately 2.0 ℃/min (room temperature → 85 ℃ ≤ 30min no-load nonlinearity).
                 
             | 
        
        
            | 
                 
                    7.
                 
             | 
            
                 
                    Cooling rate:
                 
             | 
            
                 
                    Approximately 1.2 ℃/min (room temperature → -40 ℃ ≤ 50min no-load nonlinearity).
                 
             | 
        
        
            | 
                 
                    8.
                 
             | 
            
                 
                    Indoor wind speed:
                 
             | 
            
                 
                    ≤ 1.0m/s (specialized guide layer for large laboratories).
                 
             | 
        
        
            | 
                 
                    9.
                 
             | 
            
                 
                    Test area size:
                 
             | 
            
                 
                    Width, height, and depth are 2.0 * 2.0 * 2.0m, 8.0 cubic meters, and 4 square meters.
                 
             | 
        
        
            | 
                 
                    10.
                 
             | 
            
                 
                    External dimensions:
                 
             | 
            
                 
                    Width, height, and depth are 2.2 * 2.65 * 2.2m (excluding protruding parts such as units and detachable motors).
                 
             | 
        
        
            | 
                 
                    11.
                 
             | 
            
                 
                    Power supply/total power:
                 
             | 
            
                 
                    14.5kw/380V ± 10%, 50HZ ± 5%, three-phase five wire safe grounding.
                 
             | 
        
        
            | 
                 
                    12.
                 
             | 
            
                 
                    Operating current:
                 
             | 
            
                 
                    -Approximately 21A at a constant temperature of 40 ℃.
                 
                
                    Approximately 8.0A at a constant temperature of 85 ℃.
                 
             | 
        
        
            | 
                 
                    13.
                 
             | 
            
                 
                    Test environment conditions:
                 
             | 
            
                 
                    The ambient temperature is 25 ℃, the relative humidity is ≤ 85%, and it ranges from 86 to 106 KPa.
                 
                
                    Under the condition of no sample inside the test box.
                 
             | 
        
        
            | 
                 
                    14.
                 
             | 
            
                 
                    Cooling method:
                 
             | 
            
                 
                    Standard water-cooled model, optional air-cooled model (* air-cooled model has higher noise than water-cooled model).
                 
             | 
        
        
            | 
                 
                    15.
                 
             | 
            
                 
                    Unit noise:
                 
             | 
            
                 
                    ≤ 75dB (A), measured at a distance of 1 meter directly in front of the center of the gate and a height of 1.2 (when there is no other interference or noise).
                 
             |