Number
|
Technical projects
|
Specification parameters
|
1.
|
Temperature limit range:
|
-70 ℃ to 100 ℃ (no-load).
|
2.
|
Temperature control range:
|
-40 ℃ to 85 ℃ (standard load).
|
3.
|
Temperature fluctuation:
|
≤ ± 0.5 ℃.
|
4.
|
Temperature uniformity:
|
± 2.0 ℃.
|
5.
|
Temperature deviation:
|
≤ ± 2.0 ℃.
|
6.
|
Heating rate:
|
Approximately 2.0 ℃/min (room temperature → 85 ℃ ≤ 30min no-load nonlinearity).
|
7.
|
Cooling rate:
|
Approximately 1.2 ℃/min (room temperature → -40 ℃ ≤ 50min no-load nonlinearity).
|
8.
|
Indoor wind speed:
|
≤ 1.0m/s (specialized guide layer for large laboratories).
|
9.
|
Test area size:
|
Width, height, and depth are 2.0 * 2.0 * 2.0m, 8.0 cubic meters, and 4 square meters.
|
10.
|
External dimensions:
|
Width, height, and depth are 2.2 * 2.65 * 2.2m (excluding protruding parts such as units and detachable motors).
|
11.
|
Power supply/total power:
|
14.5kw/380V ± 10%, 50HZ ± 5%, three-phase five wire safe grounding.
|
12.
|
Operating current:
|
-Approximately 21A at a constant temperature of 40 ℃.
Approximately 8.0A at a constant temperature of 85 ℃.
|
13.
|
Test environment conditions:
|
The ambient temperature is 25 ℃, the relative humidity is ≤ 85%, and it ranges from 86 to 106 KPa.
Under the condition of no sample inside the test box.
|
14.
|
Cooling method:
|
Standard water-cooled model, optional air-cooled model (* air-cooled model has higher noise than water-cooled model).
|
15.
|
Unit noise:
|
≤ 75dB (A), measured at a distance of 1 meter directly in front of the center of the gate and a height of 1.2 (when there is no other interference or noise).
|