major function
    ● Automatically calibrate the probe without the need for manual calculation;
    ● Automatically generate DAC and AVG curves using standard test blocks;
    ● Automatically display echo related parameters (depth D, horizontal P, distance S, wave amplitude H);
    16 flaw detection channels, which can freely input and store any industry's flaw detection standards;
    Can store 980 A-scan images;
    ● Four (I, V, U, X) weld seam types with graphical functions to visually determine the reflection position;
    ● Automatic calculation of defect equivalent φ Value;
    ● High sampling rate, low noise;
    ● Two probe methods: single crystal and twin crystal;
    ● Echo suppression function with warning light prompts.
    
    Auxiliary functions
    ● Gate sound and light alarm.
    ● The automatic gain function improves the efficiency of flaw detection;
    ● Two input methods: oblique probe angle and K value;
    ● Peak memory and echo envelope quickly determine the highest echo;
    ● Connect the USB interface to the PC for data transmission and automatically generate inspection reports;
    
    Product features
    The buttons are simple and easy to operate;
    The large screen highlights and can meet the requirements of flaw detection in sunlight and culverts.
Specification parameters
    
        
            | 
                 
                    project
                 
             | 
            
                 
                    technical data
                 
             | 
        
        
            | 
                 
                    Scanning range(Mm)
                 
             | 
            
                 
                    0~6000
                 
             | 
        
        
            | 
                 
                    Vertical linear error
                 
             | 
            
                 
                    ≤4%
                 
             | 
        
        
            | 
                 
                    Horizontal linear error
                 
             | 
            
                 
                    ≤0.1%
                 
             | 
        
        
            | 
                 
                    Sensitivity margin for flaw detection
                 
             | 
            
                 
                    ≥60dB(Deep)200mmφ2Flat bottomed hole
                 
             | 
        
        
            | 
                 
                    dynamic range
                 
             | 
            
                 
                    ≥32dB
                 
             | 
        
        
            | 
                 
                    resolving power
                 
             | 
            
                 
                    ≥34dB
                 
             | 
        
        
            | 
                 
                    Frequency range(MHz)
                 
             | 
            
                 
                    0.5~20
                 
             | 
        
        
            | 
                 
                     Gain adjustment(DB)
                 
             | 
            
                 
                    0~110
                 
             | 
        
        
            | 
                 
                    Material sound velocity(M/s)
                 
             | 
            
                 
                    1000~9999
                 
             | 
        
        
            | 
                 
                    Echo suppression
                 
             | 
            
                 
                    0~80%
                 
             | 
        
        
            | 
                 
                    Working temperature (℃)
                 
             | 
            
                 
                    -27~70
                 
             | 
        
        
            | 
                 
                    External dimensions(Mm)
                 
             | 
            
                 
                    240 * 160 * 40
                 
             | 
        
        
            | 
                 
                    Weight(Kg)
                 
             | 
            
                 
                    1.3
                 
             | 
        
        
            | 
                 
                    Optional configuration
                 
             | 
            
                 
                    Probe, standard test block, laptopPCCommunication softwareBNCcable
                 
             | 
        
    
    (Standard blocks are not standard, they are optional and can be selected according to the industry)
     
Standard configuration
    
        
            | 
                 
                    main engine
                 
             | 
            
                 
                    1platform
                 
             | 
        
        
            | 
                 
                    Straight probe
                 
             | 
            
                 
                    1branch
                 
             | 
        
        
            | 
                 
                    Oblique probe
                 
             | 
            
                 
                    1branch
                 
             | 
        
        
            | 
                 
                    ac adapter
                 
             | 
            
                 
                    1individual
                 
             | 
        
        
            | 
                 
                    PCCommunication software
                 
             | 
            
                 
                    1share
                 
             | 
        
        
            | 
                 
                    BNCcable
                 
             | 
            
                 
                    1root
                 
             |