                    model
                 
             | 
            
                 
                    LY-FS16BS-R400
                 
             | 
        
        
            | 
                 
                    Equipment functions
                 
             | 
            
                 
                    Applicable for conducting six waterproof level tests on products, including IPX1, IPX2, IPX3, IPX4, IPX5, and IPX6
                 
             | 
        
        
            | 
                 
                    parameter
                 
             | 
            
                 
                    Inner box size
                 
             | 
            
                 
                    Length 900 * Depth 900 * Height 1050mm
                 
             | 
        
        
            | 
                 
                    External dimensions
                 
             | 
            
                 
                    Approximately 1350 in length * 3700 in depth * 1920mm in height
                 
             | 
        
        
            | 
                 
                    Turntable diameter
                 
             | 
            
                 
                    ¥ 400mm
                 
             | 
        
        
            | 
                 
                    Maximum load-bearing capacity of the turntable
                 
             | 
            
                 
                    50KG
                 
             | 
        
        
            | 
                 
                    Turntable lifting height
                 
             | 
            
                 
                    500-800 mm
                 
             | 
        
        
            | 
                 
                    Turntable tilt
                 
             | 
            
                 
                    0 ° or 15 ° (horizontal direction)
                 
             | 
        
        
            | 
                 
                    Turntable speed
                 
             | 
            
                 
                    1-5r/min (precisely adjustable)
                 
             | 
        
        
            | 
                 
                    Rotating mode of turntable
                 
             | 
            
                 
                    Continuous forward rotation/alternating forward and reverse rotation/intermittent rotation
                 
             | 
        
        
            | 
                 
                    Flow regulation method
                 
             | 
            
                 
                    Using a flow meter with an adjustable panel to manually adjust the flow rate
                 
             | 
        
        
            | 
                 
                    Test time
                 
             | 
            
                 
                    0-999 min
                 
             | 
        
        
            | 
                 
                    Total mass of equipment
                 
             | 
            
                 
                    250kg
                 
             | 
        
        
            | 
                 
                    IPX12 parameters
                 
             | 
            
                 
                    Drip plate area
                 
             | 
            
                 
                    600 * 600mm
                 
             | 
        
        
            | 
                 
                    IPX12 spray hole size
                 
             | 
            
                 
                    0.4mm diameter
                 
             | 
        
        
            | 
                 
                    Distance between drip holes
                 
             | 
            
                 
                    20 * 20 mm
                 
             | 
        
        
            | 
                 
                    rainfall
                 
             | 
            
                 
                    IPX1=1mm/min; IPX2=3mm/min
                 
             | 
        
        
            | 
                 
                    Raindrop panel lifting method
                 
             | 
            
                 
                    Non lifting (fixed)
                 
             | 
        
        
            | 
                 
                    IPX34 parameters
                 
             | 
            
                 
                    Pendulum radius size
                 
             | 
            
                 
                    R400mm
                 
             | 
        
        
            | 
                 
                    R400 swing tube flow rate
                 
             | 
            
                 
                    IPX3=1.1L/min ± 5%; IPX4=1.8L/min ± 5%
                 
             | 
        
        
            | 
                 
                    Water spray aperture size
                 
             | 
            
                 
                    0.4mm
                 
             | 
        
        
            | 
                 
                    Water spray aperture spacing
                 
             | 
            
                 
                    50mm
                 
             | 
        
        
            | 
                 
                    Pendulum inner diameter
                 
             | 
            
                 
                    15mm
                 
             | 
        
        
            | 
                 
                    Pendulum angle
                 
             | 
            
                 
                    120 º (IPX3), 350 º (IPX4)
                 
             | 
        
        
            | 
                 
                    Swinging speed of pendulum tube
                 
             | 
            
                 
                    240 ° round-trip 4S, 720 ° round-trip 12S (precise control)
                 
             | 
        
        
            | 
                 
                    IPX56 parameters
                 
             | 
            
                 
                    IPX5 water spray aperture
                 
             | 
            
                 
                    6.3mm
                 
             | 
        
        
            | 
                 
                    IPX6 water spray aperture
                 
             | 
            
                 
                    12.5mm
                 
             | 
        
        
            | 
                 
                    IPX5 spray water flow rate
                 
             | 
            
                 
                    12.5 ± 0.625L/min
                 
             | 
        
        
            | 
                 
                    IPX6 spray water flow rate
                 
             | 
            
                 
                    100 ± 5L/min
                 
             | 
        
        
            | 
                 
                    Water pressure (Mpa)
                 
             | 
            
                 
                    Adjust water flow according to regulations
                 
             | 
        
        
            | 
                 
                    Spray distance
                 
             | 
            
                 
                    2.5 meters (distance between the nozzle and the center of the turntable)
                 
             | 
        
        
            | 
                 
                    Fixation method of spray gun
                 
             | 
            
                 
                    Mechanical fixation (program setting, adjusting spray gun angle)
                 
             | 
        
        
            | 
                 
                    Electrical parameters
                 
             | 
            
                 
                    Equipment power supply
                 
             | 
            
                 
                    380V, 50Hz
                 
             | 
        
        
            | 
                 
                    Test sample power supply
                 
             | 
            
                 
                    220V, 50Hz
                 
             | 
        
        
            | 
                 
                    Equipment power
                 
             | 
            
                 
                    3.0KW
                 
             | 
        
        
            | 
                 
                    Security protection function
                 
             | 
            
                 
                    Leakage protection, water shortage protection, short circuit protection
                 
             |